[email protected]
tel. 042-825-7983
fax. 042-825-7984

TS6B 6mm Laser Shutter

TS6B 6mm Laser Shutter

Part Number: TS6B1T0

Option 

-Blade Option

-Connector

Key Features

-6mm aperture

-Bi-stable operation

-Extremely low-profile form-factor

-RoHS Compliant

-Transfer time on opening: 1.7 milliseconds

-Total opening time: 3.1 milliseconds

-Configured for the VED24 Shutter Driver

 

문의하기

→ 상품 상세 정보 : TS6B 6mm Laser Shutter

Overview


The Uniblitz TS6B is a 6mm bi-stable laser shutter that withholds an extremely low-profile construction. The device’s lack of an external protruding actuator makes it very slim and flexible for system integration. Precision exposure control and laser switching are applications that benefit greatly from the precise, repeatable characteristics of the TS shutter series.

Bi-stable shutter devices, like the TS6B, require no power to hold the blades in either the open or closed state.

Specifications


For additional technical information please see the TS6B Laser Shutter Datasheet.


Coil Resistance Pulse Voltage to Open Hold Voltage
8 OHMS +10.7 VDC 1 N/A

 

Weight Operating Temp. Max. Opening Bounce Max. Closing Bounce Max. Freq. of Operation 2 Number of Shutter Blades
0.17 oz (4.70 g) -10 – +65 °C 5% 5% 15.0 Hz / 45.0 Hz 1

 

1 Peak voltage as provided by the VED24 Driver
2 (Continuous/Burst) Continuous frequency rating specified at shutter’s minimum exposure pulse. Burst frequency rating specified for four (4) seconds maximum with one (1) minute minimum between bursts.

 

Shutter Timing


timing-diagram-white


The question regarding enhancement of shutter speed with the application of user supplied lubricants has been repeatedly asked. It is our experience that lubricating the shutter blades will actually slow the shutter down and eventually render the shutter inoperable. Under no circumstances should any type of lubricant be applied to the shutter blade area.

 

TS6B (w/ Uniblitz driver and Teflon® coated blades) Time (msec.)
O – A:     Delay time on opening after current applied 1.4
A – C: Transfer time on opening 1.7
O – C: Total opening time 3.1
C – E: Min. dwell time with min. input pulse 9.3
B – F: Min. equivalent exp. time 11.0
D – E: Delay time on closing after current applied 1.4
E – G: Transfer time on closing 1.7
A – G: Total window time 12.7
MET: Min. exposure time 11.0
TEP: Typical exposure pulse >11.0

 

Technical Drawing


Uniblitz TS6B 6mm Laser Shutter Drawing

Go to Top