Part Number: LS3S2T0
Option
-Driver Compatibility
-Blade Option
-Electronic Sync.
-For use in Vacuum
-Mount
Key Features
-3mm aperture
-Uni-stable operation
-Laser Energy Rating: 5 W/mm2 (with “ZM” blades)
-RoHS Compliant
-Transfer time on opening: 0.5 milliseconds
-Total opening time: 1.5 milliseconds
-Configured for the VCM-D1 Shutter Driver
→ 상품 상세 정보 : LS3 3mm Laser Shutter
Overview
The Uniblitz LS3 is a high-performance laser shutter, ideal for precision exposure control, laser switching, and other applications that benefit from the precise, repeatable characteristics of the LS shutter series. These include low level laser chopping, and pulse gating, selection, and modulation to 200Hz.
Uni-stable shutter devices, like the LS3, require power to hold the blades in the open state (though the device can be custom-configured for “normally open” operation).
Specifications
For additional technical information please see the LS3 Laser Shutter Datasheet.
Coil Resistance | Pulse Voltage to Open | Hold Voltage1(Nominal) |
---|---|---|
48 OHMS | +65 VDC | +10 VDC |
Weight Housed | Operating Temp. | Max. Opening Bounce | Max. Closing Bounce | Max. Freq. of Operation2 | Number of Shutter Blades |
---|---|---|---|---|---|
7.41 oz (0.21 kg) | 0 – 80 °C | 15% | 5% | 50 Hz / 200 Hz | 2 |
1 Voltage level required across actuator coil when being held in the open position.
2 (Continuous/Burst) Continuous frequency rating specified at shutter’s minimum exposure pulse. Burst frequency rating specified for four (4) seconds maximum with one (1) minute minimum between bursts.
Electronic Synchronization System
Shutter Timing
The question regarding enhancement of shutter speed with the application of user supplied lubricants has been repeatedly asked. It is our experience that lubricating the shutter blades will actually slow the shutter down and eventually render the shutter inoperable. Under no circumstances should any type of lubricant be applied to the shutter blade area.
LS3 (w/ VCM-D1 driver and Teflon® coated blades) | Time (msec.) | |
---|---|---|
0-A: | Delay time on opening after current is applied | 1.0 |
A-C: | Transfer time on opening | 0.5 |
0-C: | Total opening time | 1.5 |
C-E: | Min. dwell time with min. input pulse | 0.8 |
B-F: | Min. equivalent exp. time | 1.5 |
E-G: | Transfer time on closing | 0.6 |
A-G: | Total window time | 1.9 |
MET: | Min. exposure time | 2.0 |
TEP: | Typical exposure pulse | >2.3 |