Fluorescence Mini Cube with 3 ports

Fluorescence Mini Cube with 3 ports

 FMC3_E(460-490)_F(500-550)

The single excitation band fiber photometry measurements use Connectorized Fluorescence Mini Cube with 3 ports, one for the excitation light, one for the fluorescence detection and one for the sample under test.

 

문의하기

→ 상품 상세 정보 : Fluorescence Mini Cube with 3 ports

The single excitation band fiber photometry measurements use Connectorized Fluorescence Mini Cube with 3 ports, one for the excitation light, one for the fluorescence detection and one for the sample under test. The cube has a dichroic mirror to separate the excitation light from the fluorescence emission and may incorporate narrow bandpass filters that limit excitation or fluorescence spectrum. Currently we offer configurations for GFP-like or RFP-like fluorophores. The 3 ports Mini Cube filters can be customized on request.

 

On the drawing E is for excitation, F for fluorescence and S is for fixed sample port.